China Dairy ›› 2025, Vol. 0 ›› Issue (10): 88-93.doi: 10.12377/1671-4393.25.10.14

• QUALITY CONTROL • Previous Articles     Next Articles

Application and Optimization of Soleris Microbiological Rapid Detection System Coliform Group Detection of Frozen Desserts

CHEN Chen, ZHANG Huiping, JIN Wei, LI Zhenxia, LI Ruixue, JIN Xiao, ZHANG Jingze   

  1. Inner Mongolia Mengniu Dairy Industrial Co.,Ltd.,Hohhot Inner Mongolia 011517
  • Online:2025-10-25 Published:2025-11-14

Abstract: [Objective] This study aimed to optimize the pretreatment steps of the microbiological rapid detection method for coliforms in frozen desserts using the Soleris microbiological rapid detection system,with the results of the traditional national standard plate counting method for coliforms as the calibration dataset,and to establish a regression calibration curve for coliforms,providing efficient technical support for microbiological detection of frozen desserts.[Method] First,the pretreatment processes for dairy-containing and non-dairy frozen desserts were optimized and unified: 25 g of melted and mixed samples were placed in 225 mL of sterile diluent,and the pH was adjusted to 6.7±0.2 if necessary.Then,the Soleris microbiological rapid detection system and the traditional national standard plate counting method (GB 4789.3—2016) were used to simultaneously detect coliforms in frozen desserts,and the applicability of the Soleris system was verified by comparing the detection results of the two methods.[Result] The detection rate of the Soleris microbiological rapid detection system for 30 samples with different pollution levels reached 100%,with no false positive or false negative results. The regression calibration curve of the detection results was y=-0.405 3x+6.061 9 (R²=0.869 6),indicating a significant negative correlation between the Soleris detection time and the number of coliforms. Repeatability verification showed that the maximum intra-group relative standard deviation was only 3.4%,indicating good system stability.[Conclusion] On the basis of maintaining the characteristic of "negative correlation between positive warning time and the number of target microorganisms",the optimized Soleris microbiological rapid detection system significantly shortens the detection cycle (warning time 7.2~11.9 h),effectively reduces the detection workload and improves the detection efficiency,which can meet the needs of rapid quality control in the production process of frozen desserts.

Key words: frozen desserts, coliforms, Soleris microbiological rapid detection system, pretreatment optimization, regression calibration curve

[1] 李丽,刘瑶,韩亚娟.发达国家乳业发展经验及对中国的启示[J].食品科学技术学报,2017,35(2):84-88.
[2] 唐铭. 基于钻石模型的中国乳业现状分析[J].商情,2017(19):58.
[3] 熊强,史纯珍,刘钊.食品微生物快速检测技术的研究进展[J].食品与机械,2009,25(5):133(184.
[4] 樊蕴秀,李敏,郑隽,等.实时光电微生物检测技术在乳品质控中的应用[J].乳业科学与技术,2013,36(5):25-28.
[5] 邹泽宇,隋青,王凤丽,等.果蔬类食品货架期的预测方法及研究进展[J].中国果菜,2025,45(2):9-17.
[6] 贺晨,孙鸿燕,邵丽筠,等.多重PCR结合基因芯片技术检测11 种致病菌方法的建立[J].中国实验诊断学,2011,15(4):587-591.
[7] 王晓宏,熊正英.多重PCR-质谱联用技术在病原体检测中的应用及比较[J].微生物学免疫学进展,2011,39(4):81-83.
[8] Park Y,Cho Y H,Jee Y,et al.Immunomagnetic separation combined with real-time reverse transcriptase PCR assays for detection of norovirus in contaminated food[J].Applied & Environmental Microbiology,2008,74(13):4226-4230.
[9] 金慧英,陶开华,李越希,等.大肠菌群O157菌体抗原基因检测芯片的制备[J].生物技术通讯,2003,14(6):279-283.
[10] 鲍俊杰,杨红玲,郭彩娇,等.酶联免疫吸附测定法与化学发光微粒子免疫分析法检测广州地区儿童HBsAb结果分析[J].中西医结合肝病杂志,2009,19(1):44-46.
[11] 谭秀山,俞雅琼,赵兴.食品微生物检测技术的研究进展[J].中国食品工业,2012(3):51-53.
[12] Cunha A F,Lage A D,Pereira e Araújo M M,et al.Evaluation of microbiological quality of UHT milk drink and UHT milk cream by ATP-Bioluminescence[J].Arquivo Brasileiro de Medicina Veterinária e Zootecnia,2013,65(2):595-600.
[13] 易敏英,李志勇.电阻抗法快速检测鲜牛奶中细菌总数[J].中国乳品工业,2001,29(3):30-31.
[14] 李春艳,霍贵成,王德国,等.ATP生物发光法快速测定生乳中微生物总数的研究[J].食品工业科技,2008,39(7):103-106.
[15] 王宁,刘宁.流式细胞术快速检测生乳中细菌总数[J].食品工业科技,2007,28(9):197-200.
[16] 孔丽娜,李祖明,吴聪明,等.生乳微生物快速检测技术研究进展[J].食品研究与开发,2013(24):268-271.
[17] 北京安普生化科技有限公司,Soleris系统在乳制品微生物快速检测领域的应用[J].食品安全导刊,2012(7):34-35.
[18] 中华人民共和国国家卫生和计划生育委员会,国家食品药品监督管理总局.GB4789.3—2016 食品安全国家标准食品微生物学检验大肠菌群计数[S].
[19] 萨日那,康俊巍,李玉英,等.快速微生物法与国标法对比实验研究[J].中国食品工业,2013(9):58-59.
[20 邢益俊,何瑛,林木娣,等.Soleris微生物快速检测系统在巴氏杀菌乳中的应用验证[J].中国乳业,2022(11):85-88,93.
[21] 张雪苍,周国磊,傅科涵,等.Soleris检测技术在焙烤食品微生物检测中的应用研究[J].食品科技,2020,45(4):321-326.
[1] XING Yijun, HE Ying, LIN Mudi, LAI Shuying, CAO Xuesi, YANG Aijun, JI Kunfa. Soleris Microbial Rapid Detection System in Application Verification of Pasteurized Milk [J]. China Dairy, 2022, 0(11): 85-89.